Single Zone Gas MonitorP/N: 3015-4256 Rev 13 253.6. Clearing the Stored PPM Log, Alarm and Fault DataUp to 200 gas measurements, and 30 alarm and fault events are stored by the monitor.To clear stored data, first display the data to be cleared by using the PPM LOG, ALARMS or FAULTSfunction (page 22). Next, press the Right Arrow Keypad and ENTER buttons at the same time. A single,long tone should be heard when the data has been successfully cleared.3.7. Reset to Factory Default SettingsIMPORTANT! Performing this function will delete all program parameters, alarms, faults,and ppm log files.With AC power removed from the monitor, press and hold down the Factory Default Switch (pages 10and 11). Apply AC power and listen for a beep and then a long tone. Release switch. At this time themonitor will purge the IR bench and auto set the IR detector voltage to its baseline 0 ppm value.After warming up, reprogram the SZ monitor to the desired settings as needed.3.8. The P-CHK Function3.8.1. OverviewThe P-CHK function (Pressure Check Function) (page 22) is useful to a service technician fortroubleshooting a flow fault problem. The monitor will trigger a flow fault if the pressure drop from ambientis less than 0.02 psi during a purge cycle, and 0.06 psi during a measurement cycle.3.8.2. Keypad FunctionsFrom the Data Display screen, use the Keypad buttons to place the arrow (>) on the display next to theP-CHK function. Then press ENTER to display the Pressure screen.The Left Arrow Keypad button toggles the purge valve open and closed. Note that an asterisk (∗)appears when the purge valve is open (purging).The Down Arrow Keypad button toggles the pump ON and OFF.Pressing the ENTER button stores the current manifold pressure shown on the left to the ambientpressure shown on the right (must be done with the pump OFF).3.8.3. Screen DisplayManifold Pressure – Current manifold pressure in psia.Stored Ambient Pressure – Stored ambient pressure in psia.Pressure Difference – The difference between the current manifold pressure and the stored ambientpressure.Fault Code – Current fault code (page 23).Purge Valve Asterisk (∗) – When viewing the Pressure screen, the purge valve can be opened andclosed by pressing the Left Arrow Keypad button. An asterisk appears on the display when the purgevalve is open.14.72 * AMB14.810.09dif <0000>Manifold Pressure ∗ Stored Ambient PressurePressure Difference Fault Code