7 Conditioning Operating Instructions BA ULTIMATE POWER + / BA ULTIMATE40 64 01 553 D3551.201.03.03.02 For further processing, BA ULTIMATE POWER + / BAULTIMATE should be dry and free of residue.3. Blow BA ULTIMATE POWER + / BA ULTIMATE out with max. 3 bar.4. Maintain mechanical parts manually [ → 43].5. Maintain the push button chuck manually [ → 45].6. Pack BA ULTIMATE POWER + / BA ULTIMATE in packagingmaterial suitable for sterilization and storage. e.g. paper/plasticcomposite packaging.7. Conduct sterilization [ → 47].7.3 Manual cleaning and disinfectionManual conditioning is possible in exceptional casesIMPORTANT: Manual conditioning is possible in exceptional cases ifthe national/regional regulations are followed. The national/regionalregulations are to be checked before.6401553.book Seite 40 Montag, 13. Juni 2016 4:27 16