©Adam Equipment Co. 2018 2913.0 ERROR CODESERRORCODESDESCRIPTION SUGGESTIONS- -oL - - Over-range Remove weight from the scale.If the problem persists contact your dealer or AdamEquipment for assistance.Err 1 Time Setting Error Enter time using correct format and reasonable values.Format: hh:mm:ssErr 2 Date Setting Error Enter date using correct format and reasonable values.Format: yy:mm:ddErr 4 Zero Setting Error The scale was outside the normal zero setting range eitherwhen it was turned on or when the [Zero] key was pressed.Remove weight from the scale and try re-zeroing again.Use the [Z/T] key to set the display to zero value.If the problem persists contact your dealer or AdamEquipment for assistance.Err 6 A/D out of range The values from the A/D converter are outside the normalrange.Remove the weight from the scale if overloaded.Make sure the pan is attached.Indicates the load cell or the electronics may be faulty.If the problem persists contact your dealer or AdamEquipment for assistance.Err 9 Check weigh limitserrorShown if the low limit is set higher than the current highlimit. Reset High limit or different low limit.FAIL Calibration error. Improper calibration (should be within + 10% of the factorycalibration). The old calibration data will be retained until thecalibration process is complete.If the problem persists contact your dealer or AdamEquipment for assistance.