14. Instant helpIn case of an error in the program process, briefly turn off the balance and disconnect from powersupply. The weighing process must then be restarted from the beginning.HelpShould other error messages occur, switch balance off and then on again. If the error messageremains inform manufacturer.30Fault Possible causeThe displayed weight does notglow.The displayed weight ispermanently changingT h e w e i g h i n g r e s u l t i sobviously incorrectWeight variation or calibrationfailure• The balance is not switched on.• Mains power supply interrupted (mains cable defective).• Power supply interrupted.• Rechargeable battery inserted incorrectly or empty• Draught/air movement• Table/floor vibrations• Weighing plate has contact with other objects.• Electromagnetic fields / static charging (choose differentlocation/switch off interfering device if possible)• The display of the balance is not at zero• Adjustment is no longer correct.• Great fluctuations in temperature.• Warm-up time was ignored.• Electromagnetic fields / static charging (choosedifferent location/switch off interfering device if• Check transport Lock is removed or not, if any• Prefer Linearity (see. chapter 9.2)