72 Testing blood glucoseFarbe: P 249 – Prüfmittelnummer: 20373Testing blood glucose4baNotesThe Accu-Chek Softclix Plus lancing device is intended only forpatient self-monitoring by a single person. The lancing devicemust not be used on more than one patient due to the risk ofinfection.Healthcare professionals testing blood glucose in a multi-patientsetting must not use the Accu-Chek Softclix Plus lancing deviceto collect blood and should therefore remove the lancing devicefrom the meter. The lancing device may be replaced with a blank-ing plate (see Chapter 10.1).Only use Accu-Chek Softclix lancets in your Accu-Chek SoftclixPlus lancing device. Using any other lancets may damage thelancing device or impair its function. This may result in injury.4.2.1 Undocking the lancing device from the meterIf you wish to use the lancing device undocked from the meter, detach itas follows:P Push the slide button in the direction indicated by the arrow (1) andhold it there.P Withdraw the lancing device in the direction indicated by the arrow(2).