Reviewing Stored ResultsA maximum of 1000 test result records are stored by the handheld, and can be reviewedby accessing the Data Review Function:1. Press to turn on handheld.2. Press to change screen to Administration Menu.3. Press (Data Review).4. At the Data Review screen, choose the category ofresults for review. Use to move from the mostrecent record to additional records.Data Review DefinitionsPatient: Records for patients are recalled by scanning or manually entering apatient ID number. If no patient ID is entered, all patient test recordsare recalled when is pressed. Not all tests may be displayed onfirst screen. Press handheld’s arrow keys to page through all screens.Control: All quality control test records.Proficiency: Proficiency testing is a type of Quality Test. Not required for testingunder a Certificate of Waiver.Cal Ver: Calibration Verification is a type of Quality Test. Not required for testingunder a Certificate of Waiver.Simulator: All external and internal Electronic Simulator records.All: All test records in the handheld’s memory.List: Records are listed with cartridge type, date and time of test, andpatient or control ID (lot number). Records can be selected for viewingor printing using the numbered keys. Pressing the number keycorresponding to a record selects the record; pressing the number keya second time deselects the record. To view one or more records,select the records and press the key.MENU2213:26 18MAY10Administration Menu1 - Analyzer Status2 - Data Review3 - Quality Tests4 - Customization5 - Set Clock6 - Transmit Data7 - UtilityData Review1 – Patient2 – Control3 – Proficiency4 – Cal Ver5 – Simulator6 – All7 – ListENT3ENT38 Art: 720672-00J, Rev. Date: 11-Oct-17