English 5SymbolsSymbols Function / Meaning TreatmentTurns on or off the deviceSets the clock / Does not store the current dataDirection guide to install batteryDirect currentSN Serial numberDate of manufactureType BF: Device, cuff and tubing aredesigned to provide special protectionagainst electric shocks.Appears when measurement is in progressand flashes when the pulse is detected.Measurement is in progress --remain as still as possible.Previous measurement stored in MEMORYAppears when the battery voltage istoo low for the device to work properly.Replace all batteries with newones.Appears if the systolic and diastolicmeasurements are within 10 mmHg ofeach other.Verify that the air hose isproperly connected and that theexhaust rate is between 2 and 5mmHg. Check for air leakage.Appears if the pressure value isunstable due to movement during themeasurement.Try the measurement again.Remain very still during themeasurement.Appears if the pressure value did notincrease during pump up.Verify that the air hose isproperly connected and that theexhaust rate is between 2 and 5mmHg. Check for air leakage.System error Remove the batteries. Installthem again properly.Appears if pulses are not detectedcorrectly.Apply the cuff properly and trythe measurement again.Attention symbol See instruction for use.SYS Systolic blood pressure in mmHgDIA Diastolic blood pressure in mmHgPUL./min Pulse per minuteEC directive medical device labelWEEE labelManufacturerREPEC EU-representativeRefer to instruction manual/bookletClass II devicePolarity of DC jack