SM500FField mountable paperless recorder 2 InstallationIM/SM500F Rev. Z 132.2.3 Pipe-Mounting (Optional)Referring to Fig. 2.6, secure the recorder to a pipe as follows:a Fit two M6 x 50mm long hexagon-head screws through one clamp plate as shown.b Using the appropriate holes to suit vertical or horizontal pipe, secure the clamp plate to thepipe-mounting bracket using the two M6 x 8mm long hexagon-head screws and two of the springlock washers.c Position the pipe mounting bracket into the recesses on the rear of the recorder as shown andsecure with the two bracket securing screws. Ensure the plastic washers remain in the positionsfitted.d Secure the recorder to the pipe using the remaining clamp plate, spring lock washers and nuts.Note. If removal of knockouts is required, refer to Section 2.3.1, page 15.Fig. 2.6 Installing the Recorder – Pipe-Mounting (Optional). 5 !