21Solution A: Enter point 1 (to 5)Deg.C and Adjust are shown alternately on the upper display line. Using the andkeys, adjust the temperature reading (in 5 increments) to the first of the temperatures onthe pH/temperature curve.pH and Adjust are shown alternately on the center display line. Using the and keys,adjust the pH reading (in 0.01pH increments) to the pH reading that corresponds to thetemperature reading entered above.Notes.1) For accurate calibration, it is important to repeat the above for buffer solution A at all 5points along the pH/temperature curve.2) The displayed temperature value increases automatically by 5C from the value set forthe previous point. The setting may be increased but not decreased.Solution B: Enter point 1 (to 5)Solution B set up is identical to solution A set up.Note. For accurate calibration, it is important to repeat the above for buffer solution B at all5 points along the pH/temperature curve.The analyzer calculates the pH/temperature relationship from the data entered.See Section 4.1.3.Alter Sec. Code not set to zero (Section 5.8) – see Section 5.1.Alter Sec. Code set to zero (Section 5.8) – see Section 5.2.…4.1 Sensor Calibration4.1.2 Set Up User Defined Buffers (pH Only)4 SETUP…A: Enter point 535.0Deg.C4.02pHSensor Cal. ASECURITY CODESet Auto BuffersBuffer Typeset to User(see Section 4.1.1)A: Enter point 115.0Deg.C0.00pHB: Enter point 535.0Deg.C4.02pHB: Enter point 115.0Deg.C3.99pHCONFIG. DISPLAYA: Enter point 115.0Deg.C3.99pH