Single and dual input analyzers for pH/Redox (ORP)AX416, AX436, AX460, AX466 & AX468 4 SetupIM/AX4PH Rev. P 214.1.2 Set Up User Defined Buffers (pH Only)Solution A: Enter point 1 (to 5)Deg.C and Adjust are shown alternately on the upper display line. Using the andkeys, adjust the temperature reading (in 5 º increments) to the first of thetemperatures on the pH/temperature curve.pH and Adjust are shown alternately on the center display line. Using the andkeys, adjust the pH reading (in 0.01 ph increments) to the pH reading that correspondsto the temperature reading entered above.Notes.1) For accurate calibration, it is important to repeat the above for buffer solution A at all5 points along the pH/temperature curve.2) The displayed temperature value increases automatically by 5 ºC from the value setfor the previous point. The setting may be increased but not decreased.Solution B: Enter point 1 (to 5)Solution B set up is identical to solution A set up.Note. For accurate calibration, it is important to repeat the above for buffer solution Bat all 5 points along the pH/temperature curve.The analyzer calculates the pH/temperature relationship from the data entered.See Section 4.1.3 page 22.Alter Sec. Code not set to zero (Section 5.8) – see Section 5.1, page 28.Alter Sec. Code set to zero (Section 5.8) – see Section 5.2, page 29. ) -* *-* +*'5522.&/3. "//0 -*' ) -*' -** ,- .+ -*