AW T 2 1 0 | 2 - W I R E C O N D U C T I V I T Y, P H/O R P, P I O N T R A N S M I T T E R | O I/A W T 2 1 0 - E N R E V. C218 Diagnostic alarmsConfigurationFailureCheck FunctionOff SpecificationMaintenance RequiredAlarm code in the format:--XPPP.NNN--X = Alarm classF = FailureC = Check functionS = Off specificationM = Maintenance requiredPPP = Alarm priority(high number = high priority)NNN = Alarm numberAlarm classAlarm message(uppercase)Alarm groupRecovery action(lowercase)Check temperaturesensor and performTC recognition--S010.047--MANUAL TEMPERATURECOMPENSATION MODEBack ExitFigure 19 Example diagnostic alarmNoteAlarms are listed in alarm priority order (high number = high priority alarm).Diagnostic message ALARM MESSAGE Recovery action pH 2-electrodeconductivity4-electrodeconductivityToroidalconductivity HART FF PAElectronics--F119.018--SENSOR MODULEMEMORY FAILURE Change sensor module Electronics--F118.023--COMMS MODULEMEMORY FAILURE Change comms module Electronics--F106.032--CURRENT OUTPUTNOT CALIBRATEDTrim outputIf problem persistschange comms module Configuration--C098.041--DATASIMULATION Configuration--C097.030--CURRENT OUTPUTFIXEDEnable loop currentmode. Disable loop test/trim& PV cal. Process--C096.031--CURRENT OUTPUTSATURATEDAdjust engineeringrange Electronics--F088.016--SENSOR MODULEFAILURE Change sensor module Process--F087.040--OPEN CABLE OR SENSOROUT OF SOLUTIONCheck sensor wiringVerify that sensoris in solution Electronics--F086.000--PRIMARY VARIABLEINPUT READ ERRORCheck sensorIf problem persistschange sensor module Electronics--F085.003--2ND PRIMARY VARIABLEINPUT READ ERRORCheck sensorIf problem persistschange sensor module Operation--M084.038--SHORTED CABLE ORGROUND LOOPS PRESENT Check sensor wiring Sensor--M083.007--SENSORPOLARIZATIONCheck processCheck sensor wiringClean sensor Table 2 Diagnostic alarms