2. Warnings and Precautions10 I PaX-i3D Smart● DO NOT place the PC or peripheral equipment connected to the PC in theimmediate vicinity of the patient.● The equipment and PC should be connected to a common protective earth.● Never overload the equipment’s circuit by sharing it with too many appliances.● Use the same power circuit for the PC and the equipment.❚ Combining this equipment with other devices● Do not connect this equipment to devices which are not part of the system.● Do not connect this equipment to a Multiple Portable Socket-Outlet (MPSO) orextension cord.❚ Electromagnetic compatibility● This equipment complies with IEC standard 60601-1-2.Medical electrical equipment is subject to special EMC preventive measures. It must be installedand operated as specified in the Installation Requirements document.If high-voltage systems, radio link systems or MRI systems are located within 5 m of the unit,please observe the specifications stated in the installation requirements.Portable and mobile RF communications equipment may interfere with medical electricalequipment. Therefore, the use of mobile wireless phones in medical offices or hospitalenvironments must be prohibited.For more details, refer to the document Electromagnetic Compatibility Information.Please also observe the ESD (ESD: Electro-Static Discharge) protective measures describedbelow.❚ Static Discharge● Connector pins or sockets bearing ESD warning labels must not be touched orinterconnected without observing ESD protective measures.Electrostatic discharge (ESD)