EN-46 PX Series BalanceTABLE 9-4. SPECIFICATIONS (continued)InCal Approval Model PX124M PX224M PX323M PX523M PX3202M PX5202M PX4201MCapacity (g) 120 220 320 520 3200 5200 4200Readability d (g) 0.0001 0.0001 0.001 0.001 0.01 0.01 0.1Verification Interval e(g) 0.001 0.001 0.01 0.01 0.1 0.1 0.1Class I I II II II II IIRepeatability(STDEV) (g) 0.0001 0.0001 0.001 0.001 0.01 0.01 0.1Linearity (g) ±0.0002 ±0.0002 ±0.002 ±0.002 ±0.02 ±0.02 ±0.2Stabilization TimeTypical (s) 4 4 2 2 1 1 1Sensitivitytemperature drift(PPM/K)±3 ±3 ±3 ±3 ±3 ±3 ±10Typical MinimumWeight USP(USP K=2,U=0.10%)200 mg 200 mg 2 g 2 g 20 g 20 g 200 gOptimized Min-Weight (g)(USP, u=0.10%, k=2)SRP ≤ 0.41d*82 mg 82 mg 0.82 g 0.82 g 8.2 g 8.2 g 82 gUnits g, mg , ct g, kg , ctApplications Basic Weighing, Parts Counting, Percent Weighing, Dynamic Weighing, DensityDeterminationPlatform Size(diameter, mm) 90 90 120 120 180 180 180Span CalibrationPoints (g) 50, 100 100, 200 150, 300 250, 500 1500,30002500,50002000,4000Linearity CalibrationPoints (g)50, 75,10050, 100,150, 200100, 200,300200, 300,400, 5001000,2000,30002000,3000,4000,50001000,2000,3000,4000Tare Range To capacity by subtractionPower Supply Power input: 100-240V ~ 200mA 50-60Hz 12-18VAPower output˖12 VDC 0.5AAssembledDimensions(W x D x H) (mm)209 x 321 x 309 209 x 321 x 98Communication RS232, USBOperatingTemperature Range 10°C to 30°CStorage TemperatureRangeHumidity: maximum relative humidity 80% for temperatures up to 30°C, decreasing linearlyto 50% relative humidity at 40°CStorage Conditions -10°C to 60°C, humidity 10% to 90%, without condensationNet Weight 10 lb / 4.5 kg 7.7 lb / 3.5 kgShipping Weight 15.4lb / 7 kg 11 lb / 5 kgShipping Dimensions(W x D x H) (mm) 507 x 387 x 531 550 x 385 x 291Note:*SRP refers to the standard deviation for n replicate weighings (nŐ10).