1-3. Make measurement of the LED currenta. Press to perform a single point measurement.b. Press to repeat single point measurements periodi-cally.c. Press to stop making measurements periodically.1-4. Turn off the channel outputa. Press to edit Channel 1 Source value, and then enter0 V to set Source value to 0 V.b. Press Ch1 to turn off Channel 1 Output relay.2. Make current – voltage sweep measurement2-1. Change View mode to Single Viewa. Press repeatedly untilSingle View for Channel 1 isshown in the display.2-2. Configure the condition to source and measurea. Press , then press to set Channel 1 V/I SourceFunction to V Source. (If can’t be found on the Assistkeys, press to change the keys.)b. Press and set Channel 1 Limit value to 100 mA.c. Press , then press to set Channel 1 Measure-ment Parameter to Current.2-3. Change View mode to Graph Viewa. Press repeatedly untilGraph View is shown in thedisplay.2-4. Perform the measurementa. Press Ch1 to turn on Channel 1 Output relay, andthen press to perform a sweep measurement. Duringthe measurement, the status will be shown.2-5. View the measurement result grapha. Press to adjust the scale of the graph after finishingthe measurement.This information is subject to change without notice.© Keysight Technologies, 2014 - 2015Published in USA, February 27, 20155991-3951ENwww.keysight.comKeysight | Making Current–Voltage Measurement Using B2901A/02A/11A/12A - Demo GuideMeasurement parametersare updated wheneverTrigger Button is pressedMeasurement parametersare updated periodicallySource ShapeSweep Sub-Panel“Auto” indicator is turned on“Auto” indicator is turned offd. Press , and then press to show SweepSub-Panel.e. Press , then press to turn on Single Linear SweepMode. After turning on Single Linear Sweep Mode, you cansee Source Shape which shows the single linear sweepmode.f. Rotate to select Channel 1 Sweep Parameters and setthem up as below. (Start: 0 V, Stop: 2 V, Points: 101, Step:20 mV)2-6. Change the graph scalea. To change the graph scaling of Y-axis from LINEAR to LOG,rotate and press to select Y-axis scaling.b. To change the graph scale value of Y-axis, rotate and pressto modify Y-axis scale value.2-7. View the list of the measurement dataa. To see the list of the measurement result, press ,then press to open Measure Result dialogue.b. Rotate and press to select List Area. Then rotateto scroll the data list.For other unlisted countries:www.keysight.com/find/contactuswww.keysight.con/find/precisionSMU